CONDUCTIVITY METER:
The meter is used on sites exposed to rain, dust, and robust handling. Basically, it comes in two parts i.e. Meter & Sensor.
Sensor:
It is the sensing element of meter, which senses the conductivity and sends the signal to the meter.
Meter:
This is the unit that processes the information received from the sensor and displays the actual conductivity in the line.
Technical Specifications:
SENSOR |
MOC |
SS316/ABS with ¾” Tee |
End Connections |
½” / ¾” BSPM |
Tee |
½” / ¾” equal tee with BSPF threading |
Cell Constant |
K=0.1(other values optional) |
Cable |
2 core shielded 3 meter length
(Higher length optional) |
Max. operating pressure |
5 kg/cm2 |
METER |
Measuring Range |
Various Ranges |
Parameter sampling rate |
2 seconds |
Resolution |
0.01 to 1 depending on range with auto ranging |
Accuracy |
± 2% of FSD |
Linearity |
1% |
Supply/Power |
230VAC ± 15% / 3 VA |
Display |
4 Digits LED 7 SEMENT |
Enclosure |
ABS weather proof |
Calibration |
Using front keypad |
Physical Dimensions |
96 x 96 x 150 mm
92 mm x 92 mm (panel cut out) |
Temperature |
0-60 degree Celsius |
Terminals |
Sensor screwed type (on pcb)
Mains – 3-core mains chord through gland |
Trouble Shooting:
TROUBLE |
PROBABLE CAUSE |
ACTION |
DURING COMMISSIONING |
Fluctuating
Reading |
Sensor not dipped in the line properly |
Loop the outlet tubing such that the tee always remains |
Wrong wiring |
Connect the wires as per drawing |
DURING NORMAL SERVICE |
Display
Shows
Erroneous
Reading |
Sensor not dipped in the line .properly |
Loop the outlet tubing such that the tee always remains flooded with water |
Deposition on the sensor |
Remove the sensor and clean it with Polish paper |
No Display |
High Voltage |
Check MOV/ FUSE, if it is burnt then replace it with new one |
Frequent
Fuse failure |
MOV short |
Replace MOV*
As well as FUSE** |
MOV* - 14 mm Dia. And 320V AC
FUSE** - 500mA |